So far we have seen how to program the 3468A
for specific operations and how
to read
back the data.
But now
it is time to move
on and
look at special
HP-IL
commands
and advanced
programming
topics.
3468A
Response to Loop Messages
The
following
topics,
arranged
in alphabetical
order,
deal
with
BASIC
language
controller HP-IL commands
and the way the 3468A
responds to them.
Refer to
the I/O programming manual of your controller for specific information on syntax
and actions taken by the HP-IL interface when sending the message. The examples
given apply to the -hp- Model 85 controller.
CLEAR
9
(device clear)
CLEAR 901
(selected device clear)
Comments
Upon
receiving the CLEAR
message,
the 3468A
will be placed into its turn-on
state. Any partially completed I/O operation is aborted. The SRQ Mask is cleared
except for the power-on
SRQ
bit (bit 7) which
is set by switch
3 on the rear
panel. The power-on state for the 3468A is: DC Volts function, Autorange, Inter-
nal Trigger, Autozero
On, and 5 1/2 digit mode.
68