Abnormal
Multimeter Readings
Message
Do This
press:
OR
ADRS
(2)
The display now shows the HP-IL ad-
dress of the 3468A.
Table 2-2. 3468A Messages
ERROR
MESSAGES
(see Self Test)
ERROR
1 - Calibration RAM error
ERROR 2 -
Internal microprocessor RAM
error
ERROR
4 - Internal microprocessor ROM
error
ERROR
8 - A/D converter error detected during Self Test
A-D ERROR - An A/D converter error was detected while trying to take a reading
represents the summation of the individual error message numbers. For example, ERROR 9 would indicate
0
that errors 1 and 8 were detected.
13