YPP6820 Explosion-Proof Loop-Powered Rate/Totalizer
General Compliance
Information

Electromagnetic Compatibility

EMC Emissions •
CFR 47 FCC Part 15 Subpart B Class A
emissions requirements (USA)
ICES-003 Information Technology
emissions requirements (Canada)
AS/NZS CISPR 11 Group 1 Class A
ISM emissions requirements
(Australia/New Zealand)
EN 55011 Group 1 Class A ISM
emissions requirements (EU)
EN 61000-6-4 Emissions requirements
for Heavy Industrial Environments -
Generic
EMC Emissions
EN 61326-1 EMC requirements for
and Immunity
Electrical equipment for measurement,
control, and laboratory use – industrial use

Product Ratings and Approvals

CSA
Explosion-proof for use in:
Class I, Division 1, Groups B, C and D
Dust-ignition-proof for use in:
Class II/III, Division 1, Groups E, F and G; T6
Flame-proof for use in:
Zone 1, Ex d IIC T6
Ta = -40 to 75°C
Enclosure: Type 4X & IP66/IP68
Certificate number: 2605742
ATEX
Explosion-proof for use in:
II 2 G D
Ex db IIC T6 Gb
Ex tb IIIC T85°C Db IP68
Ta = -40 to 75°C
Certificate number: Sira 13ATEX1121X
IECEx
Explosion-proof for use in:
Ex db IIC T6 Gb
Ex tb IIIC T85°C Db IP68
Ta = -40 to 75°C
Certificate number: IECEx SIR 13.0042X
ATEX/IECEx Specific Conditions of Use
1.
The equipment label and epoxy coating may generate
an ignition-capable level of electrostatic charges under
certain extreme conditions. The user should ensure
that the equipment is not installed in a location where
it may be subjected to external conditions (such as
high-pressure steam) which might cause a buildup of
electrostatic charges on non-conducting surfaces.
Additionally, cleaning of the equipment should be
done only with a damp cloth.
2.
Flameproof joints are not intended to be repaired.
3.
All entry closure devices shall be suitably certified as
"Ex d", "Ex t" and "IP66/68" as applicable. Suitable
thread sealing compound (non-setting, non-insulating,
non-corrosive, not solvent based, suitable for the
ambient rating) must be used at the NPT conduit
entries to achieve the IPx8 rating while maintaining the
Ex protection concept.
Year of Construction
This information is contained within the serial number with
the first four digits representing the year and month in the
YYMM format.
For European Community:
The YPP6820 must be installed in accordance with the
ATEX directive 2014/34/EU, the product certificate Sira
13ATEX1121X, and the product manual.

EU Declaration of Conformity

For shipments to the EU and UK, a Declaration of
Conformity was printed and included with the product.
For reference, a Declaration of Conformity is also
available on our website.