Motorisation technicals characteristics
Detail of the features of the Control Box:
1. Display information:
- To view the information stored in the control box, press the Up and Down
buttons on the housing for 5 seconds.
2. Factor of operation:
- Recognizes for the cylinder (Up / Down) the non-compliance with the rest time
of the cylinder.
3. Overloading cylinder # 1:
- Counts the number of times an attempt to transfer a person weighing beyond
the maximum limit reached is made. When this is the case, the cylinder will stop
and an anvil will be displayed. It will then be impossible to continue the operation
of the device.
4. Overload of cylinder # 2:
- Count the number of times an attempt is made to move or close the feet while
the feet are stuck. When this is the case, the cylinder will stop and an anvil will
be displayed. It will then be impossible to continue the operation of the device.
5. Incomplete charge:
- Counts all incomplete charges in the battery below 24.6V. This number corres-
ponds to the number of charges that damage the battery.
6. Usage time:
- Displays the cumulative usage time of the device. From 600,000 seconds of
use, the maintenance key is displayed - it is necessary to perform a complete
control of the device.
7-List of error codes:
E001: Required time setting
E002: Low battery charge level
E003: Abnormal operating time (duty factor)
E004: Cylinder overload N ° 1
E005: Cylinder overload N ° 2
E006: Maintenance required (600,000 seconds)
Installation Guide / Stand-up: WAYUP 5 Manufactured by NAUSICAA Médical S.A.S. / Approved by Ghizlane Labrosse (Biomedical