Cleaning & Maintenance
BEFORE HANDLING:
• Unplug the power cord from the mains.
• Check that all electrical components are connected to each other.
• Clean the electrical enclosures of the cylinders and wire controls if they have been splashed with
body fluids, especially urine.
DAILY / BETWEEN EACH PATIENT:
• Clean with a surface cleaner.
• Clean the electrical enclosures of the cylinders and wire controls if they have been splashed with
body fluids, especially urine.
-Clean and disinfect all parts in direct contact with the patient (handles, foam support under the
patella, etc.).
COMPLETE CLEANING AND DISINFECTION:
• Clean the entire chassis of the unit with a pre-disinfectant detergent.
• Clean the actuator with a cloth lightly moistened with a pre-disinfectant detergent solution
• Never clean the appliance, especially the electrical system, with a high-pressure cleaner, water hose
or similar.
• Disinfect the appliance (e.g. by spraying with surface disinfectant or by spraying with disinfectant
solution), please observe the contact time of the product used.
MAINTENANCE AFTER CLEANING AND DISINFECTION:
• Check that there is no visible damage to the unit.
• Check that no parts are missing.
• Check that the wheels are functioning properly and that no material is interfering with their running
(hair, carpet ends, etc.).
• Check that the controls are working properly and that the remote control and the actuator are
connected to the battery pack.
• Check the integrity of the electrical cables (cylinder and remote control).
ATTENTION:
• The detergents used should be pH neutral.
• Avoid abrasive products and solvents as they may damage the surface of the unit.
If you notice a malfunction, contact the Service Department:
Verticalizers & Lifts
Telephone: 04 66 71 71 80
Fax: 04 66 71 71 81
Mail: sav@nausicaa-medical.com
Mounting Guide / Verticalizer : BLUE WAYUP 3
25
Manufactured by NAUSICAA Médical S.A.S. / Approved by Ghizlane Labrosse (Biomedical Engineer)