A-MIP 100 Installation and user guide

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the A-MIP 100 and is the answer not in the manual?

Questions and answers

Summary of Contents for Aivia A-MIP 100

  • Page 1 A-MIP 100 Installation and user guide...
  • Page 2 THE AIVIA CONTAINS BATTERIES. THERE IS A RISK OF EXPLOSION IF THE • Leave a space of at least 13 cm around the AIVIA to BATTERIES SUPPLIED FOR USE BY THE AIVIA ARE assure proper ventilation.
  • Page 3: Table Of Contents

    Contents Installation Important Aivia presentation Identification Label Initial opening Aivia installation Installing batteries in the Aivia Setting up the defibrillator Installing seals Pictograms Operation Night mode operation Alarm operation Using the Aivia After using the defibrillator Aivia Maintenance Troubleshooting Aivia 100 W...
  • Page 4: Installation

    Aivia reference and serial number available. Initial opening • To open the Aivia door before fixing it to a wall, use a small screwdriver. • Slide the screwdriver between the door and the Aivia Electronic compartment.
  • Page 5: Aivia Installation

    Bornavie charter or authorized by the Aivia manufacturer. • If the Aivia is installed in a public street or place with public access, install a fixed ground plate (abutment) or column. Please remind that it must cumply with regulation in force regarding accessibility normative in each country and / or aerea.
  • Page 6: Installing Batteries In The Aivia

    Surface. Screw Plug.. Mounting screw, M6 minimum, M8 maximum. • Ensure that the Aivia is powered up by checking that Flat washer 14mm (0,55 in) maximum. the red warning lights are flashing. The screw head and washer together must not exceed 8 mm (0,31 in) thickness.
  • Page 7: Setting Up The Defibrillator

    • .Regularly inspect seal condition • You must install the AED status sensor on the • Seals must be installed on both sides of the AIVIA AIVIA W before establishing the seals. Please refer to the AIVIA 100 W section Seal installation •...
  • Page 8: Night Mode Operation

    • Clean the surfaces of the Aivia using a soft damp cloth. The visual warning will be given by flashing red lights. Wipe the Aivia with a soft, clean and dry cloth. Do not use If your Aivia is equipped with an audible alarm, it will chemical products or cleaners, which may damage the sound for about 3 minutes (30 seconds for an 'Aivia W).
  • Page 9: Troubleshooting

    • Red lights do not flash when the door is opened. 1.Replace batteries. 2.Contact distributor. • Siren does not sound when door is opened. 1.Check the Siren option on the identification label. 2.Replace batteries. 3.Contact distributor. •Red lights flash (Aivia 100 W). Replace batteries...
  • Page 10: Aivia 100 W

    • Connection to the AIVIANET server. Aivia 100 W After powering up, the connection to the AIVIANET The Aivia 100 W permits to verify and trace the AED server is automatic. monitoring information of the AIVIAnet serve Visit the supervision website: monitoring.aivia.eu...
  • Page 11: Specifications

    Sound level: 90 dB /1 metre. • Autonomy: Over 1 year (12 hours night) Autonomy Aivia 100 W: Over 3 years (12 hours night) • Aivia operating temperatures and humidity : Minimal temperature : : 0ºC (32°F). 40ºC (104°F). Maximal temperature:: Relative humidity: 95% without condensation.
  • Page 12 2016-11-04 *A-MIP100-EN-211* info@aivia.eu...