Technical Specifi Cations - Novis ProCair Plus User Manual

Table of Contents

Advertisement

Technical Specifi cations
SYSTEM
CONTROL UNIT
MATTRESS
DIMENSIONS
CONTROL UNIT
DIMENSIONS
OPERATING
ENVIRONMENT
22
MODEL
SYSTEM CODE
CAPACITY
NO OF CELLS
COMPLIANCE
ARTG
PART NO.
CONTROL SYSTEM
CYCLE TIME
SUPPLY VOLTAGE
MAXIMUM CURRENT
FUSE RATING
MIN / MAX PRESSURE
PROTECTION TYPE
INGRESS PROTECTION RATING
LENGTH
WIDTH
HEIGHT
WEIGHT
TOP COVER
MATERIAL
BASE COVER
AIR CELL
HEIGHT
224 mm
350 mm
WIDTH
135 mm
DEPTH
WEIGHT
3.1 kg
OPERATION
AIR HUMIDITY
STORAGE
OPERATION
AMBIENT TEMPERA-
TURE
STORAGE
ATMOSPHERIC PRESSURE RANGE
ALTITUDE
ProCair Plus
APMPC-R02
40 - 220 kg
20, including 3 static head cells, 7 narrow heel cells
All cells include a static lower chamber (cell-in-cell)
IEC60601-1, IEC60601-1-2 and IEC60601-1-11
289458
APMPC-CU02
Digital micro controller
12 minutes (fi xed)
AC100 - 240V / 50Hz-60Hz
0.3 - 0.2 A
T2AL 250V
20 ~ 60 mmHg +/- 6 mmHg
Class II Type BF
IP21
2000 mm
880 mm
200 mm
12.5 kg
PU laminated nylon
PVC laminated polyester (non-sealed base)
TPU laminated PU (sealed base)
TPU
30% to 75% non-condensing
30% to 90% non-condensing
15° C to 35° C
5° C to 60° C
700 hPa to 1060 hPa
-310 metres to 3000 metres
ProCair Plus Sealed
Base
APMPC-R02S
13 kg

Advertisement

Table of Contents
loading

This manual is also suitable for:

Apmpc-r02kApmpc-r02sk

Table of Contents