Guidance and manufactureʼs declaration - electromagnetic immunity
The Oxygen Saturation Monitor Model PULSOX-300i is intended for use in the electromagnetic
environment specified below. The customer or the user of the Model PULSOX-300i should assure
that it is used in such anenvironment.
Immunity test
Conducted RF
3 Vrms
IEC61000-4-6
150kHz to 80MHz
Radiated RF
3 V/m
IEC61000-4-3
80MHz to 2.5GHz
NOTE 1  At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2  These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a
 
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephone and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the PULSOX-300i is used exceeds
the applicable RF compliance level above, the PULSOX-300i should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the
PULSOX-300i series.
b
 
Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
IEC 60601
Compliance
Test level
level
3 Vrms
3 V/m
Electromagnetic environment -
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the PULSOX-300i, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
80MHz to 800MHz
d
1.2 P
800MHz to 2.5GHz
d
2.3 P
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance
inmeters (m).
Field strength from fixed RF
transmitters, as determined by an
electromagnetic site survey 
less than the compliance level in each
frequency range 
.
b
Interference may occur in the vicinity
of equipment marked with the following
symbol:
, should be
a
33