When influenced electromagnetically by other electronic de-
vices (e.g. near electrical appliances like TV, medial device)
When a mobile telephone is used during measurement
When influenced by abnormal hemoglobin like carbon
monoxide hemoglobin (HbCO) and methemoglobin
When pigments such as cardiogreen, intravascular dyes
and indocyanine greenare present in the blood
When finger nails are polished
Tables 1 and 2 below show the errors which may occur due to
these hemoglobins.
Difference from the SaO
<Table 1>
SaO
2
50
-0.1
70
-0.1
90
-0.2
100
-0.2
Measured value is always
lower than the actual value.
• Do not immerse the probe, the extension cable or the con-
necter of PULSOX-300i in disinfectant. Doing so may cause
a malfunction.
• When using the Extension Cable EC-300, fit and secure ca-
bles so they are not too loose or too tight.
Before connecting to a personal computer, remove the Pulsox
from the patient.
HbCO
1%
5%
10%
-0.7
-1.5
-0.7
-1.5
-0.8
-1.6
-0.8
-1.7
value to displayed value
2
<Table 2>
Methemoglobin
SaO
2
1%
50
+0.2
70
-0.6
90
-1.5
100
-1.8
Displayed value is sometimes
higher than the actual value if
SaO
is around 50%.
2
5%
10%
+1.3
+3.2
-2.3
-3.2
-6.0
-9.6
-7.5
-12.2