Wavetek 193 Instruction Manual page 16

20 mhz sweep/modulation generator
Table of Contents

Advertisement

2-4
Figure 2·6. AM Check Setup
CH1
CH2 TRIG
o
OSCILLOSCOPE
FUNC
OUT
AUX
OUT
AUX SYNC
OUT
(REAR PANEL)
Figure 2·5. FM and Sweep Check Setup
CH1 CH2 TRIG
FUNC
OUT
SYNC
OUT
AUX
OUT
MODEL 193
OSCILLOSCOPE
18
AUX GEN MODE
Set to SET FREO/AUXGEN OFF. FM center frequency of 1kHz on CH1.
19
AUX GEN FREO
FREO range to .1-3 and VERNIER (This sets the rate of modulation.)
Range, VERNIER
to 12 o'clock.
20
AUX GEN MODE
Select SET ~F and vary
Approximately 1.1 kHz.on CH1. This
and WIDTH/~F/~M
WIDTH/~F/~M for 1.1 kHz on
will
be maximum
peak
frequency
of
scope.
modulated signal.
21
AUX GEN MODE
Select FM.
Both channels: CH1 frequency varies with in-
stantaneous
amplitude
of
AUX
OUT
waveform on CH2.
22
AUX GEN FUNC
Select /'1 and full ccw rotation.
(Ramp's instantaneous amplitude controls
and SYMMETRY
sweep: A slow sweep up and fast return to
start frequency.)
23
AUX GEN MODE
Select SET START
(Ramp is held at its start voltage.)
24
FREO/START
Set dial to .02 and VERNIER
CH1: Approximately 2 Hz. (start frequency).
FREO Dial and
full ccw.
VERNIERISYM
25
AUX GEN MODE
Select SET WIDTH and rotate
CH1: Approximately 2 kHz.
and WIDTH/~F/~M
WIDTH/~F/~M cv«.
(Max sweep frequency for range).
26
AUX MODE
Select SWEEP
Sweep of main generator from 2 Hz to 2 kHz.
(If AUX GEN symmetry is rotated cw, the main
generator sweeps from 2 kHz to 2 Hz.
stepl
Control
Position/Operation
Observation
Set up the Model
193
and scope
as
shown in figure 2-5.
Table 2·1.
Checkout Procedure (coni)