2-1
2.3 ELECTRICALACCEPTANCE CHECKOUT
This checkout procedure verifies the generator opera-
tion. If a malfunction is found, refer to the Warranty in
2.2.2 Signal Connections
Use RG58U 500 coaxial cables equipped with BNC
connectors to distribute signals when connecting this
instrument to associated equipment.
Card Position
Input Vac
Fuse
100
90 to 105
3/4
amp
120
108 to 126
3/4
amp
220
198 to 231
3/8
amp
240
216 to 252
3/8
amp
1.
Disconnect the power cord at the instrument,
open fuse holder cover door and rotate fuse-pull
to left to remove the
fuse.
2.
Remove the small printed circuit board and
select operating voltage by orienting the printed
circuit board to position the desired voltage to
the top left side. Push the board firmly into its
module slot.
3.
Rotate the fuse-pull back into the normal posi-
tion and insert the correct fuse into the fuse
holder. Close the cover door.
4.
Connect the ac line cord to the mating connec-
tor at the rear of the unit and the power source.
Figure 2-1. Voltage Selector and Fuse
SEcnON2
INITIAL PREPARATION
Conversion to other input voltages requires a change
in rear panel fuse holder voltage card position and
fuse (figure 2-1) according to the following procedure.
NOTE
Unless otherwise specified at the time of
purchase, this instrument was shipped
from the factory with the power trans-
former connected for operation on
a
120
Vac line supply and with
a
%
amp fuse.
CAUTION
To prevent damage to the instrument,
check for proper match of line and instru
ment voltage and proper fuse type and
rating.
WARNING
To preclude injury or death due to shock,
the third wire earth ground must be eon-
tinuous to the facility power outlet.
Before connecting to the facility power
outlet, examine extension cords, auto-
transformers, etc., between the lnstru-
ment and the facility power outlet for a
continuous earth ground path. The earth
ground path can be identified at the
.
plug on the instrument power cord; of
the three terminals, the earth ground
terminal is the nonmatching shape,
usually cylindrical.
2.2.1 Power Connection
2.2 ELECTRICAL INSTALLATION
2.1 MECHANICAL INSTALLATION
After unpacking the instrument, visually inspect all
external parts for possible damage to connectors, sur-
face areas, etc. If damage is discovered, file a claim
with the carrier who transported the unit. The shipping
container and packing material should be saved in
case reshipment is required.