Care and maintenance

7
Care and maintenance
7.1

Maintenance instructions

The MyVeo generally requires no specific maintenance.
To ensure that it always operates safely and reliably, we recommend
that you take the precaution of contacting the responsible service
organization.
You can arrange for periodic inspections or, if appropriate, conclude
a maintenance contract with them.
Use only original spare parts for servicing.
X
Keep accessories in a dust-free place when not in use.
X
Remove dust with a pneumatic rubber pump and a soft brush.
X
Protect the devices from damp, vapors, acids, alkalis, and
X
corrosive substances.
Do not keep chemicals near the devices.
X
Protect the devices from oil and grease.
X
Never oil or grease the guide surfaces or mechanical parts.
X
To disinfect the MyVeo, use compounds from the surface
X
disinfectant group based on the following active ingredients:
• Aldehydes
• Alcohols
• Quaternary ammonium compounds
Due to potential damage to the materials, never use
products based on:
• Halogen-splitting compounds
• Strong organic acids
• Oxygen-splitting compounds.
Follow the disinfectant manufacturer's instructions.
X
• We recommend concluding a service contract with Leica
Service.
For further details refer to the cleaning, disinfection, and
sterilization instructions.
16
8
The respective applicable national laws must be observed for
disposal of the products, with the involvement of corresponding
disposal companies. The unit packaging is to be recycled.