Intended Use
This device is intended for medical purposes to illuminate body
surfaces. It is used for non-invasive visual examination of intact
skin.
This battery-operated product is designed for external exam-
ination only within professional healthcare facilities by medical
professionals.
Check the correct operation of the device before use! Do not
use the device if there are visible signs of damage.
CAUTION: Do not look directly into the LED light. Patients must
close their eyes during examinations.
In case of a serious incident with the use of this device, notify
3Gen immediately and, if required by local regulations, your
national health authority.
CAUTION: Do not use the device in fire or explosive risk area
(e.g. oxygen-rich environment).
WARNING: This product can expose you to chemicals including
methylene chloride and hexavalent chromium, which are known
to the State of California to cause cancer or reproductive toxici-
ty. For more information go to www.P65Warnings.ca.gov.
Electromagnetic Compatibility
This device complies with the EMC Emissions and Immunity
level requirements of the standard IEC 60601-1-2:2014. The
emission characteristics of this equipment make it suitable for
use in professional healthcare environment as well as residen-
tial environment (CISPR 11 Class B). This equipment offers
adequate protection to radio communication service. In the rare
event of interference to the radio communication service, the
user might need to take mitigation measures, such as relocating
or re-orienting equipment.
WARNING: Use of this equipment adjacent to or stacked with
other equipment should be avoided because it could result in
improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they
are operating normally.
WARNING: Use of accessories other than those provided by
the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immu-
nity of this equipment and result in improper operation.
WARNING: Portable RF transmitters should be used no closer
than 30 cm (12 inches) to any part of the device. Otherwise,
degradation of the performance of this equipment could result.
IMPORTANT: Before initial use, please charge the internal
lithium ion battery. See "Battery and Charging" for details.
DermLite DL1 is a mobile device-compatible dermatoscope
designed to view skin lesions with high magnification and clarity.
Four LEDs provide bright, even illumination, and a cross-po-
larized spacer provides a glare-free image by cancelling the
reflected light from the skin. A high quality, three-element lens
configuration, designed to provide excellent results with leading
mobile devices, offers superb color correction and reduced
image distortion for images rich in surface detail.
Your DL1 is equipped with a magnetic adapter called Magneti-
Connect® Clamp ("MCC"), designed for attaching DL1 to most
smartphones and tablets. To attach MCC, position the clamp
(MC) over the main camera, then secure it by tightening the
knob (KN). Attach the threaded magnetic ring and place DL1
on it.
With its cross-polarized Spacer (B) and glass contact plate (E)
installed, place DL1's contact plate directly on the skin. Applying
an immersion fluid onto the skin may improve the image. Push
the power button (C) to activate the light; push it again to turn it
off. To conserve power, your DL1 is designed to automatically
turn off after 2 minutes.
Glass Contact Plate
The unit's glass contact plate (E) may be removed for cleaning
or non-contact imaging by rotating the bayonet ring (D) count-
er-clockwise. Handle the glass faceplate with great care.
Troubleshooting
Please check www.dermlite.com for the most current trouble-
shooting information. If your device requires servicing, visit
www.dermlite.com/service or contact your local 3Gen dealer.
Care and Maintenance
CAUTION: No modification of this equipment is allowed.
Your device is designed for trouble-free operation. Repairs shall
be made only by qualified service personnel.
Cleaning
The exterior of your device, except the optical parts, may be
wiped clean with isopropyl alcohol (70% vol.) prior to use on a
patient. The lens should be treated as high-quality photographic
equipment and should be cleaned with standard lens cleaning
equipment and protected from harmful chemicals. Do not use
abrasive material on any part of the equipment or immerse the
device in liquid. Do not autoclave.
Battery and Charging
CAUTION: This device uses a special 3.7V 180mAh lithium ion
battery, which can only be purchased from 3Gen or from an
authorized 3Gen dealer. Do not under any circumstances use
another battery other than the one designed for this unit.
Prior to using the DL1, connect the wider connector of the
provided USB cable to any IEC 60950-1 compliant (5V) USB
port (e.g. of a laptop, desktop computer, or Apple charger) and
the other end to the DL1 Micro USB charging port (CP). The
charging indicator (G) will illuminate red to indicate that the DL1
is being charged. When fully charged, the indicator turns green
and the device will allow up to an hour of continuous operation.
If less than 25% battery life remains, the indicator blinks.
DermLite DL1 Basic components include:
DL1 dermatoscope (four white LEDs, three-element lens, face-
plate with 10mm reticle, battery), cross-polarized spacer with
contact plate, charging cable, MagnetiConnect Clamp adapter
Warranty: 10 years for parts and labor. The battery is warranted
for 1 year.
Disposal
This device contains electronics and a lithium-ion battery that
must be separated for disposal and may not be disposed of
with general household waste. Please observe local disposal
regulations.
Technical Description
Visit www.dermlite.com/technical or contact your local 3Gen
dealer.
ENGLISH