Installation and start-up

Note
Because the Rosemount 700XA is available in different
configurations, it is possible that not all of the instructions in this
section apply to your particular gas chromatograph (GC). In most
cases, however, to install and set up a 700XA, Emerson recommends
that you follow the instructions in the same order as they are
presented in this manual.
3.1 

Installation considerations

Before installing the gas chromatograph (GC):
1.
The GC is heavy and has a high potential of injuring personnel
or damaging equipment.
2. Ensure that the connections to the enclosure meet local
standards.
3. Use approved seals: either cable glands or conduit seals.
4. Remove any packing materials before powering up the GC.
5.
Hazardous area explosion hazard
Failure to follow this warning may result in injury or death to
personnel.
14
 WARNING
Anchor the GC solidly before making electrical connections.
Until all bolts are tight, ensure that the GC is supported to
prevent unforeseen accidents.
a. Install conduit seals within 3 in. (76 mm) of the
enclosure.
b. Seal unused openings with approved blanks (plugs).
Threads for these openings are M32 x 1.5.
 WARNING
Do not use a personal computer (PC) or printer in a
hazardous area.
Emerson provides serial and Ethernet communication links
to connect the analyzer to the PC and to connect to other
computers and printers in a safe area.