Safety Instructions - Extech Instruments EX510 User Manual

Industrial multimeter
Table of Contents

Advertisement

SAFETY INSTRUCTIONS

This meter has been designed for safe use, but must be operated with caution. The rules
listed below must be carefully followed for safe operation.
1.
NEVER apply voltage or current to the meter that exceeds the specified maximum:
Function
V DC or V AC
mA AC/DC
A AC/DC
Frequency, Resistance, Diode Test,
Continuity
2.
USE EXTREME CAUTION when working with high voltages.
3.
DO NOT measure voltage if the voltage on the "COM" input jack exceeds 600V above
earth ground.
4.
NEVER connect the meter leads across a voltage source while the function switch is in
the current, resistance, or diode mode. Doing so can damage the meter.
5.
ALWAYS discharge filter capacitors in power supplies and disconnect the power when
making resistance or diode tests.
6.
ALWAYS turn off the power and disconnect the test leads before opening the covers to
replace the fuse or batteries.
7.
NEVER operate the meter unless the back cover and the battery and fuse covers are in
place and fastened securely.
8.
If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
Input Protection Limits
1000VDC/AC rms
800mA 1000V fast acting fuse
10A 1000V fast acting fuse (20A for 30 seconds
max every 15 minutes)
1000VDC/AC rms
3
Maximum Input
EX510 V2.2 4/05/07

Advertisement

Table of Contents
loading

Table of Contents