HP 85020A Operating And Service Manual page 27

Directional bridge
Table of Contents

Advertisement

Model 85020A/B
Operation
85020B OPERATOR'S CHECK
EQUIPMENT
Sweep Oscillator
.
HP 8350A/B
RF Plug-In
.
HP 83522A/83525A/B
Scalar Network Analyzer
.
HP 8756A
10-dB Attenuator
.
HP 8491B Option 010
Calibrated 75-Ohm
Short Circuit
........
HP Part No. 1250-1530
50-Ohm to 75-Ohm Minimum Loss Pads
(2)
.
.
HP 11852A
PROCEDURE
1.
Connect equipment as shown above.
2.
Set up the test equipment as follows:
On the 8350A/B, press
[INSTR PRESET]
[SHIFT]
[CW]
[5]
[0]
[MHz]
to set the signal to 50 MHz swept CW. Press
[SWEEP TIME]
[1]
[5]
[0]
[ms].
Press
[Ul
MOD]
to turn on squarewave
modulation. On the Scalar Network Analyzer, press
[SHIFT]
and
Channel 2
[MEAS RATIO]
to turn off Channel 2.
Figure 3-3.
85020B Operators Check (1 of 2)