Operation
85020A OPERATOR'S CHECK
EQUIPMENT
Sweep Oscillator
...
HP 8350A/B
RF Plug-In
.,.
HP 83525A/B
Scalar Network Analyzer.
HP 8756A
10-dB Attenuator
.
..
HP 8491B Option 010
Calibrated Short Circuit
.
HP 1153.2A
Shielded Open Circuit
....
HP Part No. 85032-60001
PROCEDURE
1.
Connect equipment as shown above.
2.
Set up the test equipment as follows:
On
the
83
50A/B,
press
[INSTR PRESET]
[SHIFT]
[CW]
[5]
[0]
[MHz]
to set the signal to 50 MHz swept CW. Press
[SWEEP TIME]
[1]
[5]
[0]
[ms].
Press
[LflMOD]
to turn on squarewave
modulation. On the Scalar Network Analyzer, press
[SHIFT]
and
Channel 2
[MEAS RATIO]
to turn off Channel 2.
Figure 3-2.
85020A Operator's Check (1 of 2)