4.2 Insertion
4.2
Insertion
Holding and manipulating the endoscope
Ch.4
Figure 4.1
1
Hold the control section of the endoscope using your left hand.
2
Operate the remote switches 1 to 4 using the left index finger.
3
Operate the UP/DOWN angulation control lever using the left thumb.
4
Maintain the insertion section using the right hand.
Insertion of the endoscope
38
WARNING
• The shape and size of the nasal cavity and its suitability for transnasal insertion
may vary from patient to patient. No endoscope, including this one, can always be
inserted transnasally into all patients. Before proceeding, always be sure to confirm
that transnasal insertion is possible with the patient by considering both the size of
the patient's nasal cavity as well as the size of the endoscope's insertion section.
Otherwise, patient injury can result or the endoscope could become lodged and
difficult to withdraw.
• Transnasal insertion of the endoscope should be performed carefully. If resistance
to insertion is felt, or the patient reports pain, stop the insertion immediately.
Otherwise, patient injury can result or the endoscope could become lodged and
difficult to withdraw.
ENF-V3 OPERATION MANUAL