Table 1-2. 8024B Specifications (cont)
AC VOLTS (Average Sensing, RMS Calibrated Sinewave):
RANGE
|RESOLUTION
BECEAAY
45 Hz to 1 kHz | 1 kHz to 2 KHz | 2 kHz to 5 kHz
200 mV
100 UV
+(0.75%
+(1.5% of
+(5% of
750V
1V
+(1% of
Not
Not
reading +2
Specified
Specified
digits)
Response Time.............
Less than 2 seconds
Overload Protection
........
750V rms or 1000V peak continuous, except
200 mV ac ranges
(15 seconds
maximum
above 300V rms).
Common Mode Rejection
Ratio (1 kQQunbalance) .....
>60 dB at 50 Hz and 60 Hz
Volt-Hz Product ............
10' max (200V at 50 kHz)
Input Impedance
...........
10 MQ in parallel with <C 100 pF
DC CURRENT
ACCURACY
BURDEN
RANGE
RESOLUTION
10UA
0.25V max
200 mA
100WLA
Response Time.............
Less than 1 second
+(0.75% of
reading +1
digit)
Overload Protection
........
2A/250V fuse in series with 3A/600V fuse.