Delta UNOKIT2 User Manual
Delta UNOKIT2 User Manual

Delta UNOKIT2 User Manual

2nd generation uno sensor calibration system

Advertisement

Quick Links

Delta Electronics, Inc.
UNOKIT2
2
nd
Generation UNO Sensor Calibration System
JERRY.HSIAO 蕭令彥
DELTA ELECTRONICS, INC.
User Manual v1.0
0

Advertisement

loading

Summary of Contents for Delta UNOKIT2

  • Page 1 Delta Electronics, Inc. UNOKIT2 Generation UNO Sensor Calibration System User Manual v1.0 JERRY.HSIAO 蕭令彥 DELTA ELECTRONICS, INC.
  • Page 2 [Revision History] Date Version Description 2024/01/05 Release the first version.
  • Page 3 [Overview] UNOKIT is the official calibration tool for the UNO Sensor (multi-in-one air quality detector) devices. It allows for the calibration of sensor module readings on the test UNO Sensor device by comparing them with the readings of the UNO Sensor benchmark, ensuring the accuracy of the sensor readings.
  • Page 4 [Contents] Storage Bag Power Supply: AC100V-240V Calibration Chamber Power Switch Safety Clasp * Ethernet Port for Engineering Use Power Cord * USB Port: MOUSE / USB DISK Benchmark Installation * Display LCD Test Device (DUT) Installation * * The power cord is recommended to be stored inside the calibration chamber when not in use. * The safety clasp can be used to quickly detach terminal connectors.
  • Page 5 [Installation Guide] Counter-clockwise release the safety clasp and open the calibration chamber. Ensure all dip switches on the Benchmark are down, then install it on the left and the test device (DUT) on the right. Connect the UNOnext with the 14-pin cable and the UNOlite with the 5-pin cable. Close the chamber to complete the installation.
  • Page 6 Connect the power cord, and then connect the USB mouse. * Please prepare your own USB mouse. Power on. * The benchmark does not need to be removed after installation unless there is a need for replacement. * After installing the benchmark, please go to [System Settings] and update the [Time Zone] and [Benchmark Calibration Date] for the first-time use.
  • Page 7 [User Interface Guide] UNOKIT App Version Connection Status and Sensor Values of UNOKIT Status Benchmark and Test Device Current Date Message Box UI Language Settings Buttons of Calibrate / Reset /Open Report System Settings...
  • Page 8 [Calibration Process] Follow the installation guide to set up the benchmark, test device and mouse and power UNOKIT on. After the system starts up, the UNOKIT status will display 'Standby,' and the serial numbers of the benchmark and test device and sensor readings of mounted sensors should both be displayed correctly. * If there is an error message, please follow the instructions.
  • Page 9 After the sensor readings mounted on benchmark and test device are displayed (not --), click on [Calibrate], and UNOKIT will start the calibration process. UNOKIT will perform the following actions: [Warmup] → [Analyze] → [Calibrate] → [Finish]. The message block will also be keeping update according to the progress. Note: If the sensor readings persistently do not display, it may indicate that the benchmark or test device lacks this...
  • Page 10 If any errors occur during the process, the message block will display a red warning. Please follow the message instructions to resolve the issue, then click [Reset] / [Next] or reboot UNOKIT. Note: In some case, errors are discovered after [Finish] stage, a fail-calibration report also be generated. After calibrating the test machine, you can refer to the [Installation Guide] to replace it with the next one and click [Reset] or [Next].
  • Page 11 [Retrieve Report] Click on [Open Report Folder ...], and the screen will display the report storage directory. All calibration [Calibration Date]_[SN of Test Device]_[PASSED/FAILED].pdf reports are PDF file, the filename format is . e.g., 20231215_2321L0170099_PASSED.pdf. Insert the USB flash disk into [USB DISK], and the LCD will display a permission allow window. Select [Open File Manager], click [OK], and the USB flash disk directory will appear.
  • Page 12 Move the PDF file from the report directory to the USB flash disk directory. Select the PDF file, right-click, choose [Cut], and then go to the flash disk directory, right-click, and select [Paste]. Click the [Eject USB Drive] button and eject the USB flash disk.
  • Page 13 [System Settings (UNOKit Mangement)] Benchmark Calibration Date: If a benchmark replacement is performed, please update the calibration date ◼ according to the information of the new benchmark. 1. Choose Date 2. Click to Update Time Zone: Please set the time zone correctly to ensure the accuracy of the report date. (e.g. +0800 ➔ ◼...
  • Page 14 System Update: When UNOKIT is connected to the internet and can access the external network, it can ◼ perform system updates. * This operation should be carried out as instructed by an engineer.
  • Page 15 [Common Errors] Error Description alert_check_fail Calibrate fail, please re-calibrate alert_cmd_fail Unknown failure, please re-calibrate alert_dut_decline Sensor offset is too large, reset or replace it alert_dut_err Sensor error, reboot or replace it fatal_alert_bad_env Bad environment for calibration, change place and then recalibrate fatal_alert_bench_err Benchmark sensor error, please reboot or replace Benchmark fatal_alert_bench_offline...