Hp- Il Interconnections Toption 003); 2-29. Channel B Input Fuse; 2-31. Operating Environment; Storage And Shipment - HP 5384A Operating And Service Manual

Frequency counters
Table of Contents

Advertisement

··~
HP 5384A and HP 5385A
SYSTEMS
a
lnstallatoon
"Talk Only" or
"addressable~',
and selects the HP-1 8 ad-
dress. Instructions for changing the address are pro-
vided in Section Ill of this manual.
2-23.
HP-18 Description
2-24.
A description of the HP-18 is provided in
Section Ill of this manual. A study of this information
is necessary if the user is not familiar with the HP-18
concept. Additional information concerning the de-
sign criteria and operation of the bus is available in
IEEE Standard 488-1978, titled "IEEE Standard Digital
Interface for Programmable Instrumentation" .
.
""""""''""
#Ill:~
SYSTEMS
The above symbol when located in the upper
corner of a page indicates HP-18 information is
contained on that page. This information may be
operation, performance, adjustments, or service
related.
2-25.
HP-IL Interconnections (Option 003)
2-26.
The HP 5384A and HP 5385A with Option 003
are compatible with the Hewlett-Packard Interface
loop. The interface loop connection is made by a pair
of two-wire balanced line cables. The cable is limited
to 10 metres (32.8 feet) for unshielded cables (from
one device to the next) and to 100 metres (328 feet) for
shielded cables. A typical interface connection is
shown in Figure 2-4.
Table 2-2. HP-IL Cable Descriptions
HP Model
Length
82167A
.5
metres
(1.64
feet)
821678
1
metre
(3.28
feet)
821670
10
metres
(32.8
feet)
2-27.
HP-IL Description
2-28.
An introductory description on HP-Il is pro-
vided in Section Ill of this manual. As HP-Il is a rela-
tively new concept, it is recommended that users un-
familiar with the Interface loop concept review this
material prior to operating the instrument remotely.
The above symbol when located in the upper
corner of a page indicates HP-Il information is
contained on that page. This information may be
operation, performance, adjustment, or service
related.
2-4
PRINTER
LISTENER
FREQUENCY
COUNTER
TALKER
LISTENER
ADDRESS 4
Figure 2-4. Typical HP-IL Interface Connection
2-29. Channel B Input Fuse
2-30.
Figure 2-5 shows the details of how to change
the fuse in the channel 8 input of the HP 5385A.
2-31. Operating Environment
2-32.
TEMPERATURE. The counter may be operated
in temperatures from 0°C to +50°C.
2-33.
HUMIDITY. The counter may typically be
operated in environments with humidity up to 95% at
40°C. However, it should be protected from extreme
temperatures which may cause condensation in the
instrument.
2-34.
ALTITUDE. The counter may be typically
operated at altitudes up to 4,600 metres (15,000 feet).
2-35.
STORAGE AND SHIPMENT
2-36. Environment
2-37.
The instrument may be stored or shipped in
environments within the following limits:
TEMPERATURE .......... -40°C to +75°C
HUMIDITY
Up to 95% noncondensing
ALTITUDE ..... 15,240 Metres (50,000 feet)