PROBLEM SOLVING

Error Occurrence and Actions

Error
Cause
・ When the user's impedance is out
Impedance is out of
measurement range.
of the measurement range
・ In case of poor contact with the
electrode
・ When the impedance is lower
or higher than the range
・ Measurement range: 100~950Ω
・ When the user's percentage of
The percentage of
body fat is out of
body fat analysis result is out of
the measurement
the allowable range
range.
・ Personal data input error
・ When the percentage of body fat
is out of range
・ If the user's height or age is outside
When the
measurement does
the input range
not proceed to the
・ Height input error
next step
・ Age input error
・ The user's weight is out of the
When the weight
cannot be measured
measurement range.
・ Weight measurement error
・ Continue moving during
measurement
・ In case of internal communication
Not connected to the
body fat analyzer.
error
54
_Problem Solving
Action
・ Measure again after cleaning the
measurement area (electrodes, hands, feet)
・ Measure again after correcting the electrode-
holding method and measuring posture.
・ Do not move your body during
measurement to avoid momentarily poor
contact with the electrode.
・ If the same error is displayed on the LCD
even during re-measurement, contact us or a
designated vendor.
・ Measure again after wiping the hand
electrode clean
・ If the same error is displayed on the screen
even during re-measurement, contact us or
a designated vendor.
・ Please enter your height and age correctly.
However, you cannot enter it when it is out
of the analysis range.
・ Measure the weight again.
However, it is not possible to measure the
user's weight if it is out of the measurement
range.
・ Turn the power off and then on.
・ If the same message is displayed repeatedly,
please contact us or a designated vendor.
Error Code Occurrence & Actions
Error code
Cause
・ Electrode and measuring parts are
38001
not detected or dirty
・ Faulty in impedance measuring
・ There is an error in measured
38003
impedance by external influence.
・ Out of measurement range of
38004
impedance
・ Range: 100~950Ω
・ Out of measurement range of PBF
38005
・ range: Under 75%
・ Continuous error for 3 times
38006
・ Disconnect Color board and Main
38007
board
User Manual BC380
Action
・ Try to measure again after cleaning the
electrode holders with soft gauze
・ Try again after cleaning the hands and soles
・ Check if foreign materials are between
electrodes and measuring parts
・ Try again in correct posture and hold the
electrodes according to the measuring method
・ Check installation condition (AC cord or
around equipment
・ Contact SELVAS Healthcare or distributor
・ Try again after cleaning the hands and soles
・ Check if foreign materials are between
electrodes and measuring parts
・ It can't measure if the impedance is out of
range
・ Try again after cleaning the hands and soles
・ It can't measure if the PBF is out of range
・ Step off and try again
・ Contact SELVAS Healthcare or distributor
・ Contact SELVAS Healthcare or distributor