2.3 CHARGING THE BATTERY
The power used by the unit comes from the internal battery, which also keeps certain
program parameters in memory even when the unit is switched off. The unit comes with
Its battery partialiy charged and it is therefore advisable to charge the battery before use.
batteries in the instrument begin to run down, the message
BATTLO appears periodically on the display. If the battery voltage goes too low the
instrument switches itself off to prevent destructive discharge of the batteries and loss of
data stored in the memory registers.
When BATTLO appears on the display, the battery should be recharged as soon as
connector at the rear of the unit. Connect the
charger to the 120 VAC, making sure that the line voltage is correct. Charge the battery
for approximately 12 hours; 20 hours if used during charging.
2.4 WIRING CONNECTIONS
The CL521 connection terminals are found on top of the unit. Connections are easily
[3?,
®''® '^®Pt
0^ the way of the display and keypad. When using the
carrying case, the flap in front of the connection panel may be shut during
use. This reduces changes in reference temperature of thermocouple temperature
measurements, thus ensuring measurement accuracy in environments where ambient
temperature may vary greatly.
The terminal panel contains two separate types of connections.
79-
®®®<^ f®'' thermocouple and millivolt measurement and
measurement/calibration switch is in the
MtAbUKtMENT position, this connection measures inputs. The CALIBRATION
^ calibration output signal. The resistance of a circuit connected to the
I c and mV connection may also be measured by using the TCii measurement function.
The Ohm and PtIOO connection is used to measure the resistance of PtIOO or other
resistive probes. Both three and four-wire connections are possible.
Used in four wire
connections