/ or damage to your equipment. The following labels identify information that is especially important to note: Note: Provides you with information that is important to successfully setup and use the SP-001. Caution or Warning: Tells you about the risk of electrical shock.
We’ve integrated an advanced suite of IIoT centric featured electronics called Smart Core. These features enable plug and play connectivity, fully featured alarms and notifications, data assurance, and data storage of up to 1499 sensor readings. Included with Your SP-001 / SP-002 •...
The optical chart below indicates the nominal target spot diameter at any given distance from the sensing head and assumes 90% energy. Figure 3 reflects the SP-001 with a field of view of 6:1. Figure 4 reflects the SP-002 with a field of view of 10:1.
The sensor should be mounted so that the measured spot size is smaller than the target. Note: A mounting bracket for the SP-001 / SP-002 is sold separately. Ambient Temperature The sensor is designed to operate in ambient temperatures from -40°C to 85°C (-40°F to 185°F).
The SP-001 / SP-002 requires a Smart Interface (IF-001-0) or Wireless Transmitter to connect to your computer. Locate the position of the keyway as a guide on the SP-001 / SP-002 prior to making the connection. See Figure 6. Figure 6 The sensor can be mounted on brackets, clamps or cutouts of your own design or you can use the fixed and adjustable Omega mounting bracket accessory: SP-001-MB.
Figure 10 Communication Interface • Connection Type: Select the type of connection you have between your SP-001 / SP-002 and your computer. USB connection is displayed in Figure 9. USB Serial connection is displayed in Figure 10. • Command Timeout: The maximum time (in milliseconds) for a command to be completed before the command is aborted.
Emissivity Dropdown and click Apply Settings to finalize. See Figure 11. Setting Emissivity via Secondary Temperature Device Note: Ensure the SP-001 / SP-002 is connected to Omega Sync and pointed at the material being measured at a distance that covers the entire field of view. Step 1: Click Calibration.
Step 3: Once the target temperature has been determined, enter the target temperature into Omega Sync. See Figure 13. Figure 13 Step 4: Click Calibrate to set the emissivity changed and exit out of the calibration pop-up. Click Apply Settings to finalize. Note: When you click Calibrate, the emissivity value in Figure 11 will change according to the value of the Target Surface Temperature entered.
Alarm Setting The SP-001 offers software adjustable alarms for your smart sensor. To begin alarm setup, follow these instructions. Step 1: Click the icon next to the input you would like to customize. Figure 15 Step 2: Customize alarm settings to your satisfaction. See Figure 16.
Accuracy Over Entire Ambient Operating Range: 4°C (3.6 °F) or 3% of reading, whichever is greater Repeatability: 1°C (1.8°F) or 1% of reading Field of View: SP-001 is 6:1 SP-002 is 10:1 Spectral Response: 5.5 to 14 microns Emissivity: 0.01 to 1.00, software adjustable...
Page 17
Where Do I Find Everything I Need for Process Measurement and Control? OMEGA…Of Course! Shop online at omega.com TEMPERATURE M U Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies M U Wire: Thermocouple, RTD & Thermistor M U Calibrators & Ice Point References M U Recorders, Controllers &...