• Periodically inspect the fluid path for air. If air is seen, stop infusion. Air infusion could result
in under or non-delivery of medication. Air embolism could result in death or serious injury to the
patient.
• Always validate infusion protocol values by checking all parameters, prior to start the
infusion. Wrong infusion protocol could result in death or serious injury to the patient.
11 APR 2014 11:45
STOP
PUMP ON HOLD
HISTORY
PROGRAM
PROTOCOL

6.5 Changing the battery and continue the same infusion

5 NOV 2012 12:45
STOP
PUMP ON HOLD
HISTORY
PROTOCOL
PROGRAM
Turn the pump OFF and change the battery, as explained in Chapter 2, paragraph 2.3.
"Power Supply" section 2.3.1 "For pumps with 9v alkaline battery". Turn the pump ON and
repeat the procedure as explained in Chapter 6, paragraph 6.4 "Resuming the same infusion
with the same patient after power off of the pump" in order to resume the infusion after OFF.

6.6 Changing the Protocol

• We recommend the default codes of the pump to be changed (through the configuration
menu) in order to secure the access to different actions related to each code.
• The supervisor of the infusion should decide to whom each code and its related actions
should be distributed. Incorrect pump programming could result in death or serious injury to
the patient.
11 APR 2014 11:45
STOP
PUMP ON HOLD
HISTORY
PROGRAM
PROTOCOL
11 APR 2014 11:45
STOP
PUMP ON HOLD
PROGRAM
HISTORY
PROTOCOL
CODE:
0000
11 APR 2014 11:45
Titr
PUMP ON HOLD
PROGRAM
CLINIC. BO
PROTOCOL
56
!
Warning
Press twice
to resume the infusion.
The programmed parameters and their perspective values
of the running protocol can be reviewed at once using the
key. History can be reviewed using
PROTOCOL
Put the pump on hold by pressing
code to stop infusion, if code protected stop of infusion is
enabled. Press
!
Warning
Put the pump on hold by pressing
code to stop infusion, if code protected stop of infusion is
enabled, please refer to Chapter 9 paragraph 9.2 "Configuration
Menu". Press
.
Press
.
PROGRAM
Enter your "programming" code and press
Select
.
PROGRAM
twice and enter the
.
twice and enter the
.
the
HISTORY
.