Warnings
• Ensure that the battery placed in the pump is in good condition and the remaining
battery power is enough for the infusion that is going to start. If the battery is not likely
to finish the infusion, replace it before infusion starts. Loss of power may result in non-
delivery of medication.
• Always inspect the battery compartment for fluid or debris before inserting the
batteries. Fluid or debris in the battery compartment may damage the battery terminals
and could result in loss of power and non delivery of medication.
• Always inspect the battery terminals for debris, grime and damages. If debris or
grime exists, clean the terminals according to cleaning instructions. If the debris or grime
persists or damage exists remove the device from service. Debris, grime or damage of
the battery contacts could result in loss of power and non delivery of medication.
• Always have a new battery available for replacement to ensure that the pump can
continue its operation for the time specified.
• Do not use rechargeable NiCd or NiMH batteries. Do not use carbon zinc batteries.
• Only accessories supplied from Micrel shall be used with the pump. Use of non-
specified accessories may impair the operation of the pump.
• Ensure that IP Connect Pack if used is ON to supply power to the pump.
• Do not pull out the bolus handset at an angle, bolus may be damaged.
• Do not remove the bolus handset by pulling the cable, bolus may be damaged.
• Do not rotate the bolus connector while removing the bolus handset.
• Do not use other external power adaptor than the ones supplied by Micrel.
• Do not plug the Micrel power adaptor to the Mains with wet / moist hands.
• The pump should be disconnected from the patient prior to being connected to a
computer.
• Do not perform uploading procedure in the presence of a patient.
• Do not connect cords supplied from Micrel to other devices rather than the ones that
are intended to be used with.
• Εxternal power adaptor should be used only indoors.
• The external power adaptor should no longer be used if the housing or cable becomes
damaged.
• Ensure that the external power adaptor is not covered when in operation, is not in
close proximity to a heating source and is not exposed to direct sunlight.
• The external power adaptor must not come into contact with liquids.
• Ensure that the pump is OFF and disconnected from any power supply before
commencing the cleaning / disinfection procedure.