5. Programming procedure
• We recommend the default codes of the pump to be changed (through the configuration
menu) in order to secure the access to different actions related to each code.
• The supervisor of the infusion should decide to whom each code and its related actions
should be distributed. Incorrect pump programming could result in death or serious injury
to the patient.
• Always inspect the Rythmic™ Evolution pump and its accessories for damaged plastic
parts, wires and presence of liquids on its surface, prior to use. If the pump or accessories
are dropped, hit, subjected to excessive moisture and/ or beyond temperature limits,
immediately take it out of use and return it to authorized service department for inspection.
5.1 Turn the pump ON/OFF
To turn the pump ON:
and check that there are no missing dots on the dark display and hear that the buzzer
sounds twice.
Keep the
Rythmic™ volution
S/N:234567890123/V1.0A
4JAN 2013 11:54:02
To turn the pump OFF:
pump "ON HOLD" during infusion press
Rythmic™ volution
OFF
www.micrelmed.com
5.2 Programming the pump
The user can program the pump by using the following options:
• By selecting and programming an existing (built-in) delivery mode
• By selecting a pre-configured protocol (through the protocol library)
36
!
Warning
Push and hold down the
held down and check that date and time displayed are correct, if
not refer to the chapter 9 "Configuration". Pump's serial
number and Software revision number is also displayed on
this screen. Release the
The pump must be "ON HOLD" in order to turn off. To put
stage just keep pressing
HOLD" screen.
Push and hold down the
appears.
. Keep the
.
twice to stop infusion, in any other
until you reach the "ON
until the OFF screen
held down