Magstim Model 200
2
10.2 Al\tIBIENT TEMPERATURE
Permissible Environment Conditions for Transport
Ambient Temperature Range:
Relative Humidity Range:
Atmosphere Pressure Range:
Storage and Operating Conditions:
Operating Temperature Range:
Storage Temperature Range:
Coil Temperature Range:
-25°C to 60°C
10% to 80% Non Condensing
50kPa to 106kPa
5°C to 35°C
-25°C to 60°C
5°C to 40°C
10.3 CAPACITOR LIFE EXPECTANCY
Life expectancy:
10.4OUTPUT
2 x 10
6
discharges at 70% power level
8 x 10
5
discharges at 80% power level
4 x 10
5
discharges at 90% power level
2 x 10
5
discharges at 100% power level
Repetition rate of stimulus:
Magnetic Pield:
Once every 2 seconds at up to O - 49% power level
Once every 3 seconds at up to 50 - 79% power level
Once every 4 seconds at up to 80 - 100% power level
Please contact Magstim Company for further information.
Maximum number of stimulations:
The maximum number of stimulations that can be achieved with a
Magstim 200
2
at an ambient temperature of 20°C, with an interval
between discharges of 4 seconds, before the stimulating coil
exceeds its permitted operating temperature. In the case of the
Double 70mm Remote Coil
(PIN
3190-00) the stimulating coil is
capable of 60, 110 and 340 stimuli at 100%, 80% and 50% power
level respectively.
Pulse Characteristics:
80µs approximate rise time, lms duration
The Magstim Company Limited
3001-23-USA
29