Voltage Selection And Fuse Rating - MAGSTIM 2002 Manual

With double 70mm remote coil
Hide thumbs Also See for 2002:
Table of Contents

Advertisement

Magstim Model 200
2
SECTION 9: l\tlAINTENANCE AND SERVICING
9.1 VOLTAGE SELECTION AND FUSE RA TING
The Magstim Model 200 may be operated from supplies in the range 11 O-
l 20V and 220-240V. The voltage selector on the rear panel and the
adjacent fuses must correspond to the supply voltage as indicated in the
table below:
Supply
Set Voltage
Fuses
Voltage
Selector To
Qty
Rating
110-120V -
120V
2
lOA T *
220-240V -
240V
2
SAT*
* T denotes Timed, or Antisurge, fuse. The use of fast acting fuses is not
recommended.
Voltage Selection Instructions
1.
2.
lMPORTANT-Ensure that the Magstim Model 200
2
is disconnected
from the mains supply before changing the fuse, or setting the voltage
selector.
Insert the tip of a small blade screwdriver, or similar tool, into slot no.
2, shown in diagram above. Gently lever up the retaining lip of the
fuse holder. The fuse holder will then slide out.
Using a pair of narrow nosed pliers, grip the voltage selector unit on
one of the metal connection plates and remove it from the Power
Entry Module.
Orient the voltage selector unit, such that the desired voltage is facing
out from the Power Entry Module.
Re-insert the voltage selector unit into the Power Entry Module.
Replace the fuse holder, ensuring that the retaining lip is properly
home, and verify that the voltage window (1) displays the correct
voltage value.
The Magstim Company Limited
3001-23-USA
24
August 24, 2006

Advertisement

Table of Contents
loading

This manual is also suitable for:

3001-013190-00