SECTION 5: PREPARATIONS FOR USE
IMPORTANT
To avoid interference problems the Magstim Model 200
2
,
and its
accessories, should not be used in the vicinity of any equipment that does
not comply with EMC Standard EN 60601-1-2.
Before using the instrument, ensure that the correct fuses are fitted and that
the voltage selector is set to suit the supply voltage (see Section 9). Check
the Magstim Model 200
2
and its accessories for any signs of damage.
1f
the
Magstim Model 200
2
,
or any of its accessories, are damaged in any way
they must not be used.
If a number of Magstim 200
2
units are being used, or the Magstim 200
2
is
part of a BiStim system, it may be desirable for the units to be stacked on
top of each other. The Magstim 2()()2 enclosure has been designed such that
the units may be safely stacked on top of each other. The maximum
number of units that can be safelv stacked on top of the base Magstim
200
2
is 3. Attempts to exceed this limit could result in crush damage to the
base Magstim 200
2
In addition, the stacked units could become unstable
and overbalance, causing injury to the patient, or user, and damage to the
Magstim units.
1. Inse1t the pneumatic connector of the foot switch to the rear panel
FOOTSWITCH socket on the Magstim Model 200
2
.
2.
If
a recording device is being used, synchronise the Magstim with the
recording instrument by following the instructions in Section 4 -
ISOLATED TRIGGER 1/0 Port.
3. Ensure that the locating spot on the stimulating coil plug is correctly
lined up with the COIL OUTPUT SOCKET, located on the front panel of
the Magstim Model 200
2
Insert the connector of the stimulating coil into
the output socket. When the stimulating coil plug is fully engaged in the
socket, lock the plug in place by turning the black locking ring clockwise
until it clicks into place.
4. Set the MAINS POWER switch on the rear of the unit to the ON (I)
position. The Unit Power Status Indicator LED will flash at approximately
0.5Hz, indicating that the unit is in a standby condition.
5. Push the ON/OFF button on the front panel of the unit. The Unit Power
Status Indicator LED will be continuously lit, indicating that the unit is
operational and is in its safe inactive default condition.