SECTION 8: SAFETY FEATURES
8.1 COIL TEMPERATURE PROTECTION
1. An indication of the stimulating coil temperature is given by the Coil
Temperature LEDs located on the User Interface Controller - see
section 3, items 6 - 8.
2. If the stimulating coil surface temperature reaches its set limit, the
REPLACE COIL indicator will illuminate and the Magstim Model
200
2
will be put into its safe inactive default condition. The default
condition indicator will illuminate.
The normal operation of the equipment can then be resumed by
allowing the stimulating coil to cool, or alternatively, by replacing the
stimulating coil with a cool one.
8.2 COIL DISCONNECTION
If
the stimulating coil is disconnected from the unit while the unit is
charged, the REPLACE COIL indicator will illuminate and the unit will
automatically select the default mode and discharge internally.
The ARl\.fED indicator will extinguish, indicating that the unit is
discharged.
8.3 OTHER SAFETY l\1EASURES
If
the instrument has not been triggered for over 1 minute after the
READY symbol is illuminated the unit will automatically select the
default mode and discharge internally
Power charge up or charge down responds instantly to the OUTPUT
CONTROL knob, minimising the waiting period or accidental
discharge at a setting higher than required.
The Magstim Model 200
2
is capable of recognising automatically the
type of stimulating coil being used, and adjusts its maximum output
power accordingly.
High quality pneumatic foot switch.