ActiGraph LEAP Instructions For Use Manual page 27

Table of Contents

Advertisement

Regulatory Information
Regulatory Symbols
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally.
WARNING: Use of accessories, transducers, and cables other than those specified or provided by
ActiGraph could result in increased electromagnetic emissions or decreased electromagnetic immunity of
this equipment or decreased electromagnetic immunity of this equipment and result in improper operation.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the ActiGraph
LEAP™, including cables specified by the manufacturer. Otherwise, degradation of the performance
of this equipment could result.
WARNING: Both the ActiGraph Leap device and charger contain magnets. Under certain conditions,
magnets and electromagnetic fields might interfere with medical devices. For example, implanted
pacemakers and defibrillators might contain sensors that respond to magnets when in close contact.
To avoid any potential interactions with these types of medical devices, keep this device and charger
a safe distance away from your medical device (more than 6 inches / 15 cm apart). Consult with your
physician and your medical-device manufacturer for specific guidelines.
If you suspect that your ActiGraph Leap device or charger is interfering with your medical device, stop
using your ActiGraph Leap device and consult your physician and your medical-device manufacturer.
CAUTION: Do not simultaneously wear and charge, service, or provide any maintenance on the product.
CAUTION: Transport and storage of this product outside of the temperature range of -20° C to 55°
C, 15 - 90%, non-condensing 700 hPa to 1060 hPa could lead to dangerous conditions.
CAUTION: Operating this product outside of the temperature range of 0° C to 40° C could lead to
dangerous conditions.
CAUTION: Modification to ActiGraph products are not permitted and will void all warranties if tampered
and/or modified. Do not modify the product in any way as this can be unsafe to the end user.
CAUTION: Do not swallow any part of this product. If a piece has been swallowed, contact your local
poison control hotline or seek medical attention as soon as possible.
NOTICE: No precautions need to be taken in the event of changes in the performance of the product.
NOTICE: Product does not have any contraindication(s)
NOTICE: Only use ActiGraph's approved charging dock to charge the device. Do not connect the
device to anything else but a computer and an ActiGraph approved product or accessory.
NOTICE: Only use ActiGraph's approved USB cables with the charging dock.
NOTICE: The ActiGraph LEAP™ may be used in any instance where quantifiable analysis of physical
motion is desirable.
NOTICE: ME Equipment charged via certified external power supply with input rated 90-264 Vac and
output of up to 5Vdc.
ACTIGRAPH LEAP™ | E.200.6025 | RELEASED: 10/09/2023 | REV: 0
|