Temperature Compensation - Xylem WTW Cond 3310 Operating Manual

Conductivity meter
Hide thumbs Also See for WTW Cond 3310:
Table of Contents

Advertisement

Operation
Criteria for a stable
measured value
28
2
Using <ENTER>, activate the Stability control function manu-
ally.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized.
Note
You can prematurely terminate the Stability control function manually
with <ENTER> at any time. If the Stability control function is prema-
turely terminated, the current measurement data are output to the
interface without the AutoRead info.
3
Using <ENTER>, start a further measurement with stability
control.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back
to the previous indication.
The Stability control function checks whether the measured values are
stable within the monitored time interval.
Measured
Time interval
parameter
Conductivity ϰ
10 seconds
Temperature
15 seconds
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.

4.4.3 Temperature compensation

The calculation of the temperature compensation is based on the pre-
set reference temperature, 20 °C or 25 °C. It appears on the display as
Tr20 or Tr25.
You can select one of the following temperature compensation meth-
ods:
Cond 3310
Stability in the time interval
Δ : better than 1.0% of mea-
sured value
Δ : better than 0.5 °C
ba75803e08
10/2017