SENSIT Technologies SPOD C1D2 Operation Manual

SENSIT Technologies SPOD C1D2 Operation Manual

Voc emissions & air pollutant monitor

Advertisement

Quick Links

937-00000-XY - OPERATION MANUAL
This operating manual is for the SPOD C1D2, both normal (937-00000-20)
and high sensitivity (937-00000-21) models.
851 Transport Drive • Valparaiso, IN 46383 (USA)
Phone: 219.465.2700 • www.GasLeakSensors.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SPOD C1D2 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SENSIT Technologies SPOD C1D2

  • Page 1 937-00000-XY - OPERATION MANUAL This operating manual is for the SPOD C1D2, both normal (937-00000-20) and high sensitivity (937-00000-21) models. 851 Transport Drive • Valparaiso, IN 46383 (USA) Phone: 219.465.2700 • www.GasLeakSensors.com...
  • Page 2: Warnings

    WARNINGS ⚠ EXPLOSION HAZARD – DO NOT CONNECT OR DISCONNECT THE CABLES TO THE SPOD C1D2 UNLESS IN A NON-FLAMMABLE AREA, AND WHEN THE INSTRUMENT IS SHUT OFF (I.E., NO LIGHT CAN BE SEEN FROM THE LED ON THE SWITCH) ⚠...
  • Page 3: Table Of Contents

    CONTENTS WARNINGS ..................2 GENERAL INFORMATION ..............4 DEVICE OVERVIEW ....................4 CONNECTION INFORMATION .............. 5 SENSOR INFORMATION ............... 6 CELLULAR SPECIFICATIONS ..............7 DEPLOYMENT GUIDE ................8 SENSOR REPLACEMENT ................ 9 NOTES ....................10 750-00088-C1D2 – Manual for 937-00000-20 and 937-00000-21. Rev. 1.0 [7/17/2023] 3...
  • Page 4: General Information

    GENERAL INFORMATION The SENSIT SPOD measures VOC levels and transmits the collected data to a remote server. It is intended for deployment outdoors and in wet locations (e.g., rain). DEVICE OVERVIEW OVERVIEW PARAMETER Weight Base unit: 6.75 lbs Fully assembled with anemometer and antenna Dimensions •...
  • Page 5: Connection Information

    CONNECTION INFORMATION The SPOD C1D2 only has a single connection port, located at the bottom of the unit, next to the sensor. This connection is for charging and direct data communications to the SPOD C1D2 unit. The port only takes connectors in the Phoenix Contact SAC-HZ-5P family of products (UL File E359524). When the port is in use, the retained locking clip must be used with the connection to prevent accidental disconnection.
  • Page 6: Sensor Information

    SENSOR INFORMATION ⚠ EXPLOSION HAZARD – ONLY SENSOR PART NUMBER 375-PID1BIXF-01 (ION SCIENCE MP3SBLBXBU2) AND 375-PID1BIHS-01 (ION SCIENCE MP3SHLHSBU2) CAN BE USED WITH 937-00000-20 AND 937-00000-21, RESPECTIVELY. OVERVIEW PARAMETER Default PID Detection Range 0.1– 30 ppm for normal version (937-00000-20) 10 ppb –...
  • Page 7: Cellular Specifications

    CELLULAR SPECIFICATIONS OVERVIEW PARAMETER Network Technology 4G/2G Carrier AT&T, T-mobile Transport Layer TCP Internet Layer IP Application Layer HTTP and MQTT Data Transfer Method HTTP POST or MQTT Topics HTTP Content Type application/x-www-form-urlencoded HTTP Body Field Identifiers &ID, &MODULE, &STAT, &DATA MQTT Content Type JSON MQTT Tags “deviceId”, “time”, “iodb”...
  • Page 8: Deployment Guide

    Power Port ⚠ EXPLOSION HAZARD – DO NOT CONNECT OR DISCONNECT THE CABLES TO THE SPOD C1D2 UNLESS IN A NON-FLAMMABLE AREA, AND WHEN THE INSTRUMENT IS SHUT OFF (I.E., NO LIGHT CAN BE SEEN FROM THE LED ON THE SWITCH) 5).
  • Page 9: Sensor Replacement

    SENSOR REPLACEMENT ⚠ EXPLOSION HAZARD – PID SENSOR REPLACEMENT SHOULD BE COMPLETED IN A NON-FLAMMABLE, CLEAN, AND DRY ENVIRONMENT. 1). Make sure the SPOD is not powered on (i.e., there is no light coming from the LED on the switch) 2).
  • Page 10: Notes

    NOTES __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ 750-00088-C1D2 – Manual for 937-00000-20 and 937-00000-21. Rev. 1.0 [7/17/2023] 10...
  • Page 11 NOTES __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ __________________________________________________________________________________________ 750-00088-C1D2 – Manual for 937-00000-20 and 937-00000-21. Rev. 1.0 [7/17/2023] 11...
  • Page 12 851 Transport Drive Valparaiso, IN 46383-8432 Phone: 219.465.2700 Toll Free: 888.4.SENSIT (473.6748) Fax: 219.465.2701 Email: info@gasleaksensors.com Website: www.GasLeakSensors.com MADE IN THE USA WITH GLOBALLY SOURCED COMPONENTS SPOD Operation Manual for 937-00000-20 & 937-00000-21 SENSIT ® Part Number: 750-00088 Version 1.0: 7/1/2023...

This manual is also suitable for:

937-00000-20937-00000-21

Table of Contents