FCML 412 N(-S), FCML 412-M12
FCML 412 N
FCML 412 N-S
FCML 412-M12(-x)
ba77197e02
01/2024
7.4

Electrical data

 2-wire connection via permanently mounted cable AK-CL 298
 12 ... 30 V DC power supply by the measuring transmitter
 Output signal 4 ... 20 mA
 4-wire connection via AK-CL cable by means of screwed connection/plug
connection
 +/- 6 ... 12 V DC power supply by the measuring transmitter
 Nominal slope: approx. -100 mV per mg/l chlorine (0 ... -1500 mV)
 5-pin M12 flange connector: for the ADA CL/IDS
adapter by means of screw/plug connection
 Power supply: +/- 5 ... 15 V DC through the MIQ/IDS module
 Nominal slope: -1000 mV/ppm
Technical data