The 8082A is a 250 MHz dual channel pulse
source with variable leading and trailing edge transition
times as fast as Ins. It also has variable pulse frequency,
delay, width, offset and amplitude. The
ment reiationship
and the polarity of either output can
be reversed. Single pulse, double pulse and square wave
are available. There are also four trigger
1 - 3
ates as a self-contained pulse source with
of the p&e parameters from the front panel controls.
trigger output frequencies are determined by the fre-
quency of an externaity applied signal. The other
parameters are varied from the front panei controis.

GENERAL INFORMATION

Gate Mode. In this mode a gating signal
enables the pulse and trigger outputs.
External Width Mode. in this mode the pulse
1 - 6
frequency and width are determined by the frequency
and width of an externally applied signal. The delay be
tween input and output is fixed. The trigger output
is the shaped trigger input signal.
ECL OUTPUT
1 - 7
The 8082A has an ECL position on each of
1 - 8
its amplitude range switches. When
the switches are se? to this position, both
automatically deliver a fixed voltage swing of
logic.