54
9.4
Demolition and disposal
The materials of the bed consist essentially of:
▪ painted or galvanized ferritic steel;
▪ plastic material in abs;
▪ elastomers;
Disassemble the bed separating the individual pieces according to the material with
which they are made, it is mandatory to dispose of the different materials in
accordance with the regulations of the country in which the bed must be removed.
Regarding the disposal of consumer products, behave as follows.
Products for the sanitization
▪ The products used for sanitation must not be discharged into urban canalizations.
▪ Inquire about the provisions in force on disposal arrangements at the local
authorities.
Batteries
▪ The battery of the engine control unit must be regularly replaced by an electrician. Used
batteries must not be disposed of with common waste, but must be delivered to the
appropriate disposal centers.
9.5

BATTERY

It is equipped with a battery for operation in case it is not possible to connect to the
power supply.
The charging time is about 10-12 hours with completely discharged battery.
It is advisable to make sure that the batteries do not drain completely, but
recharge them frequently in order to achieve a longer service life.
Batteries must be replaced, depending on use, at least after 3 years.
To optimise the life of the batteries, the control units must be connected to the network
as much as possible. Batteries must be recharged at least every 3 months. Otherwise
they are damaged and self-discharge.
When the bed is powered only by the extra battery you can perform 20 complete
movements.
When the battery reaches 50% of its charge each movement emits an audible alarm.
An operation test is recommended at least once a year.
Batteries shall only be replaced by models having the mechanical and electrical
characteristics of the following types:
▪ KOBE 1.2-6 (6V, 1.2 Ah)
▪ YUASA 1.2-6 (6V,1.2 Ah)
▪ PBQ 1.2-12 (12V, 1.2ah)
▪ KOBE1.2-12 (12V,1.2ah)
Make sure the ventilation hole is intact. The ventilation hole allows the exit of the battery
gases but does not allow the penetration of water.
WARNING: The battery has no capacity to ensure the operation of the bed for long
periods but only serves as a buffer in case of temporary power failure. Connect the bed
to the mains as soon as possible.