34951A Scpi Programming Examples; Level Mode; Waveform Mode - Keysight Technologies 34951A User Manual

4-channel isolated d/a converter with converter with waveform memory module
Hide thumbs Also See for 34951A:
Table of Contents

Advertisement

34951A SCPI Programming Examples

34951A SCPI Programming Examples
The programming examples below provide you with SCPI command
examples to use for actions specific to the DAC module.
The slot and channel addressing scheme used in these examples follow 
the form sccc where s is the mainframe slot number (1 through 8) and ccc
is the three- digit channel number. Valid channels for this module 
are 1- 4. For information on specific configurations, refer to the simplified
schematic on
For complete information on the SCPI commands used to program the
34980A, refer to the Agilent 34980A Programmer's Reference contained on
the 34980A Product Reference CD. For example programs, also refer to the
34980A Product Reference CD.

Level Mode

Example: Outputting a DC voltage level
level for the specified DAC channels. After setting the desired level, send
the OUTPut:STATe command to close the corresponding output relay and
enable outputs from the specified channels. The following command
outputs +2.5 V DC on DAC channels 1 and 2 for a module in slot 4.
SOURce:VOLTage 2.5,(@4001,4002)
OUTPut:STATe ON,(@4001,4002)
Example: Outputting a current level This command sets the output current
level on the specified channels on the DAC module. After setting the
desired level, send the OUTPut:STATe command to close the
corresponding output relay and enable outputs from the specified
channels. The following command outputs +5 mA on DAC channels 1 and
2 for a module in slot 4 and closes the output relay.
SOURce:CURRent 5E-3,(@4001,4002)
OUTPut:STATe ON,(@4001,4002)

Waveform Mode

Example: Downloading a waveform to memory and outputting waveform from DACs
The following command segment downloads a 1000- point sine waveform to
memory on the module in slot 4 and outputs the waveform from DAC
channels 1 and 2. The trace name is "TEST_SINE".
TRACe:FUNCtion 4,SINusoid, TEST_SINE, 1000
SOURce:FUNCtion:TRACe TEST_SINE,(@4001,4002)
OUTPut:STATe ON,(@4001,4002)
SOURe:FUNCtion:ENABle ON,(@4001,4002)
4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
page
8.
Agilent 34951A 4-Channel Isolated D/A Converter Module User's Guide
This command sets the output voltage

Advertisement