Do not use your eQCM 15M or any other apparatus if you think it could be hazardous. Have it checked by
qualified service personnel.
Environmental Limits
There are environmental-limit conditions on the storage, shipping, and operation of this equipment. The
eQCM 15M is not designed for outdoor use.
Storage
Ambient Temperature
Relative Humidity
Shipping
Ambient Temperature
Acceleration
Operation
Ambient Temperature
Relative Humidity
Warning:
may enter the chassis, or water vapor may condense within the chassis. Operation of an eQCM 15M that
has water within the chassis can create a safety hazard, which could result in personal injury or death.
Cleaning
Disconnect the eQCM 15M from all power sources prior to cleaning.
Use a cloth lightly dampened with either clean water or water containing a mild detergent to clean the outside
of the eQCM 15M enclosure. Alternatively, you can use isopropyl alcohol. Do not use a wet rag or allow fluid
to enter the eQCM 15M enclosure. Do not immerse the eQCM 15M in any type of cleaning fluid (including
water). Do not use any abrasive cleaners.
Service
Your eQCM 15M Electrochemical Quartz Crystal Microbalance has no user-serviceable parts inside. Refer all
service to a qualified service technician.
Warning:
Always remove the power connection before opening the eQCM 15M case.
Emission Warning
Your eQCM 15M generates, uses, and can radiate radio-frequency energy. The radiated levels are low enough
that the eQCM 15M should not create an interference problem in most industrial laboratory environments.
The eQCM 15M has been tested for radiated RF interference and has been found to be in compliance with
EN 61000-4-3.
Safety Considerations
This instrument is not designed for operation in conditions where liquid water
Never operate eQCM 15M with any cover or panel on the chassis open.
−25C to 75C
Maximum 90% non-condensing
−25C to 75C
Maximum 30 G
0C to 45C
Maximum 90% non-condensing
9