Table 4-1 shows the input ranges and resolutions supported by each MIO X Series device.
Table 4-1. MIO X Series Device Input Range and Nominal Resolution
MIO X Series Device
NI 632x/6341/6343
NI 6345/6351/6353/6355/
6361/6363/6365/6375

Working Voltage Range

On most MIO X Series devices, the PGIA operates normally by amplifying signals of interest
while rejecting common-mode signals under the following three conditions:
The common-mode voltage (V cm), which is equivalent to subtracting AI <0..x> GND
from AI <0..x>-, must be less than ±10 V. This Vcm is a constant for all range selections.
The signal voltage (Vs), which is equivalent to subtracting AI <0..x>+ from AI <0..x>-,
must be less than or equal to the range selection of the given channel. If Vs is greater than
the range selected, the signal clips and information are lost.
The total working voltage of the positive input, which is equivalent to (Vcm + Vs), or
subtracting AI GND from AI <0..x>+, must be less than ±11 V.
If any of these conditions are exceeded, the input voltage is clamped until the fault condition is
removed.
Input Range
-10 V to 10 V
-5 V to 5 V
-1 V to 1 V
-200 mV to 200 mV
-10 V to 10 V
-5 V to 5 V
-2 V to 2 V
-1 V to 1 V
-500 mV to 500 mV
-200 mV to 200 mV
-100 mV to 100 mV
X Series User Manual
Nominal Resolution
Assuming 5% Over Range
320 μV
160 μV
32 μV
6.4 μV
320 μV
160 μV
64 μV
32 μV
16 μV
6.4 μV
3.2 μV
© National Instruments | 4-3