DISASSEMBLING THE SCOPEMETER
6-3
4.
The bottom cover assembly is secured to the top cover with two M3 Torx screws (item 29) that
are accessible in the battery compartment. Use a Torx screwdriver to remove the two screws.
5.
Lift the bottom cover a little from the top cover and unfold the ScopeMeter.
NOTE: Do not damage the black gaskets and keep them with the front cover and the lower case half.
A correctly fitted gasket assures proper sealing of the ScopeMeter.
BATTERY
COVER
~~_LlQUID
CRYSTAL
DISPLAY
~
BOTTOM COVER
8T6031
Figure 6.2
Opening the ScopeMeter