KERN KFP V30 Platform Version 1.2 2023-09 Installation Instructions Contents General hints ....................3 Technical data ....................3 Technical data of the weighing cell ..................4 Basic Information (General) ................5 Documentation ......................... 5 Proper use ..........................5 Improper Use ..........................5 Warranty ............................
1 General hints These installation instructions contain all data necessary for placing and commissioning the following platforms: KERN TKFP 15V30M-A KERN TKFP 30V30SM-A, TKFP 30V30M-A KERN TKFP 60V30M-A, TKFP 60V30LM-A, TKFP 60V30XLM-A KERN TKFP 150V30SM-A, TKFP 150V30M-A, TKFP 150V30LM-A KERN TKFP 300V30M-A 2 Technical data Item no./ Type Weighi...
3 Basic Information (General) 3.1 Documentation These installation instructions contain all data necessary for placing and commissioning the platform KERN TKFP V30-A. In combination with a display unit, described below as weighing system, for operation configuration, please refer to the operating instructions of the display unit. 3.2 Proper use The platform purchased by you is intended to define the weighing value of weighed goods.
Information is available on KERN’s home page (www.kern-sohn.com with regard to the monitoring of weighing system test substances and the test weights required for this. In KERN's accredited DKD calibration laboratory test weights and weighing systems may be calibrated (return to the national standard) fast and at moderate cost. 4 Basic Safety Precautions 4.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and...
6 Unpacking, Installation and Commissioning 6.1 Installation Site, Location of Use The platforms are designed in a way that reliable weighing results are achieved in common conditions of use. You will work accurately and fast if you select the right location for your weighing system.
6.2 Unpacking, Scope of delivery Remove the underconstruction and the weighing plate from the package, remove packaging material and place the devices at the planned workplace. Check if all parts of the scope of delivery are present and free of damage. Scope of delivery: •...
6.3.1 Transport Securing General hints for transport securing • Please consult the Technical Data whether your balance is equipped with a transport securing, chap. 2. • There are different types of transport securings in KERN balances. Please see the indications explained in this chapter. •...
Page 10
Variant 1: Note on transport securing • 1 x hexagon screw on the upper side Type of transport securing with nut • 1 x threaded pin with inner hexagon on the lower side with nut 1. Remove the weighing plate. 2.
Page 11
7. Remove the hexagon screw on the upper side. Variant 2: Note on transport securing • 4 x cylinder head screw with inner Type of transport securing hexagon on the upper side 1. Remove the weighing plate. 2. Remove the transport securings. TKFP-V30-A-IA-e-2312...
6.4 Levelling Accurate weighing results require a platform with perfect horizontal alignment. During initial installation and after each change of work area it is necessary to level the platform. As the air bubble is located under the weighing plate, remove it.
8 Operation limits The platform is designed so robust that a casual excess of the maximum weighing load will not cause any damage. The static carrying capacity, i.e. the maximum admitted load depends on the type of load carrier (Position A – C). The maximum static carrying capacity must not be exceeded.
9 Servicing, maintenance, disposal Before any maintenance, cleaning and repair work disconnect the appliance from the operating voltage. 9.1 Cleaning Clean the platform with a soft cloth soaked with a mild cleaning agent. Remove the weighing plate and wipe-off the dirt and the foreign bodies under it. Do not use hard or sharp objects for this.
10 Instant help for troubleshooting In case of an error in the program process, briefly turn off the balance and disconnect from power supply. The weighing process must then be restarted from the beginning. Help: Fault Possible cause • Draught/air movement The displayed weight is permanently changing •...
11 Preload, Deadload and Overload settings Deadload** (kg) Centric overload Off-centre overload Minimum capactiy of the Maximum capacity of Item no./ Type **= already applied protection (kg) protection (kg) load cell E (kg) the load cell E (kg) preload TKFP 15V30M-A TKFP 30V30SM-A TKFP 30V30M-A TKFP 60V30M-A...
Page 17
Item no./ Type Dimensions of Load cell type Test certificte no. Accur Ratio of min. Maximum admitted Allowable Ratio of min. Error the platform load cell number of interval temperature dead load output percentage (mm) class verification sizes range return interval ( °C) ( °C)