General Information. P5; Introduction. P5 - BeaconMedaes MAT-S Installation,Operation And Maintenance Instruction

Medical gas supply system cylinder manifold system and manifold changeover system
Table of Contents

Advertisement

Medical Gas Supply System
Do not attempt to verify the pressure relief valve, under
any circumstances, by altering the regulator. Pressure
relief valve must be removed and tested off site by a
registered test centre for a certificate of conformity
issued.
Figure 1 - Manifold Changeover System (MAT-S).
1.
General Information.
28
29
10
9
8
7
6
5
4
3
2
1
22
27
26
NOTE - Isolation valves item 3, 5 and 12 are shown in
their open position, isolation valve item 14 is shown in
its closed position. This is typically the normal operating
condition for the manual valves.
30
12
11
31
1.1
Introduction.
The BeaconMedaes Medical Gas Supply System,
mainly consists of cylinder manifold system, manifold
changeover s ystem, h eader & t ailpipe a nd o ther n ecessary
components, is intended to work in conjunction with
other devices within the medical gas pipeline system
(MGPS) to assist with maintaining continuity of supply
throughout the hospital to local connection points for
eventual introduction into the human body. The device
controls the gas pressure within the specification as
required by the MGPS. In all cases other devices outside
of this scope are connected downstream of the MGPS
are required to control the final gas pressure and flow
requirements for safe supply to the patient. Therefore,
this device is not intended to control supply parameter
hazards directly to the patient, only to the MGPS.
The device including Cylinder Manifold System and
Manifold Changeover System (abbreviate as manifold
below) is principally designed for use as a primary or
secondary source of supply, or for emergency backup.
The Medical Gas Supply System supplies one of the
following medical gases to a piped distribution system,
Oxygen, Nitrous Oxide, O2/N2O 50%: 50%, Medical Air,
Surgical Air, Nitrogen & Carbon Dioxide.
See figure 1 for general arrangement and figure 2 for the
schematic diagram.
The Manifold Changeover System consists of:
13
No. Description
1
1
Stage Pressure Regulator
st
14
2
Intermediate Pressure Gauge
15
3
Intermediate Isolation Valve (Three Way)
4
2
Stage Pressure Regulator
nd
16
5
Line Isolation Valve
6
Smart Box
17
7
Touch Screen Controller
18
8
Line Relief Valve
19
9
Line Pressure Relief Exhaust Connection point
(28 mm OD Copper Tube).
20
10
Vent Valve
11
Service Valve
21
12
Pipeline Connection Point (22 mm OD Copper
Tube)
23
13
Lockable Isolation Valve
14
Non-Return Valve
24
15
Test Point Isolation Valve
25
16
Service Relief Valve
17
Terminal Unit
18
Line Pressure Gauge
19
Line Pressure Sensor
5
4109992485.00