BeaconMedaes MAT-S Installation,Operation And Maintenance Instruction page 15

Medical gas supply system cylinder manifold system and manifold changeover system
Table of Contents

Advertisement

Medical Gas Supply System
Figure 10 - Set Line Pressure.
NOTE - Line pressure will normally be pre-set from the
factory. Options are for nominal 4, 8 & 11 bar supply. This
setting will not affect the actual supply pressure, it is for
setting the default line pressure alarm. The line pressure
setting is changed by adjusting the regulator knob.
1.
Press the line pressure value and use the keyboard to
change the number.
Figure 11 - Set Pressure Units.
NOTE - Changing the pressure units will automatically
convert all pressure related control variable to the newly
selected unit. Options are Bar (default), kPa or PSI.
1.
Press the icon
unit of measure.
2.
Press the downward arrow
to change the unit.
Figure 12 - Cylinder Set-Up.
NOTE - Minimum requirement in the cylinder set-up is
to set the correct maximum pressure. This will correctly
scale the graphic on the display. The size and number
of cylinders per bank is for calculating the average gas
usage only.
1.
Press the system icon
2.
Press the Cylinder Pressure and Cylinder Volume to modify
the value by the keyboards.
3.
Press the NO. Cyl (Left) and NO. Cyl (Right) to modify the
quantity keyboards.
NOTE - The typical average gas usage is only for gas phase
in the cylinders. This calculated value will not convert
correctly for liquid phase in the cylinders. Leaving the
No Cyl/Bank setting as '0' will hide the calculated flow on
main page.
15
in the setting page.
4109992485.00