Medical Gas Supply System
Figure 13 - Line Pressure Alarm Set-up.
NOTE - Default Line Pressure Alarm setting will be automatically defined when the Nominal Line Pressure is selected.
See table 4 below for default values. These values can be manually adjusted as required to suit on site requirements.
1.
Press the system icon
2.
Press the Low Line Pressure Alarm and High Line Pressure
Alarm and modify the value by keyboard.
Table 4: Default Line Pressure Alarm Settings, and
Adjustment Range.
Default Values
Nominal
Low
Line
Pressure
Pressure
Alarm
4 Bar
3.7
8 Bar
6.4
11 Bar
9.1
NOTE - Low line pressure set-point may need to be reduced if the
device is used as a backup manifold. See table 6 in section 4 for
reference.
4109992485.00
in the settings page.
Adjustment Range
Low
High
Pressure
Pressure
Alarm
Min
Max
5.0
2.7
4.0
8.8
5.5
8.0
11.8
7.6
11.0
Figure 14 - Alarm Output Option.
1.
Press the Units of Measure
2.
Press the downward arrow
output option.
High
Pressure
Min
Max
4.0
6.0
8.0
12.0
11.0
16.5