BRUEL & KJAER 4914 Instruction Manual page 16

Rhino-larynx stroboscope
Table of Contents

Advertisement

Fig. 3.6. Connection of the Light Cable AE 6100 using the coupler
DB 2632
.
For operation at a flash frequency of 100 Hz (mean value), the average illuminance ob-
tained is as indicated in Table 3.1.
llluminance [k lux]
Type of Light
(at distance from end of optical cable)
25mm
50mm
100mm
Halogen light
240
80
20
Stroboscope light
1
60
20
5
Light Exposure [k lux s]
Photo flash
2
24
8
2
1 Mean value at
100
Hz
T00361GBO
2 Each shot 80µs (half peak intensity duration)
These values are reduced if optical cables other than the B&K standard cable are used. For exposure
calculations consider reductions from additional optical systems
Table 3.1. Average illuminance provided by Type 4914 plus Light Ca-
ble AE 6100 at the flash frequency of 100 Hz (mean value)
For use with other types of optic cable a Blank Coupler DP 0594 is supplied. The dimen-
sions of this Coupler are shown in Fig. 3.7., and when drilled with an appropriate sized
centre hole it can accommodate cables up to 17,5 mm in diameter. For optimal light
output it is important that the input face of the optic cable protrudes just under 40 mm
from the flange of the Coupler. See Fig. 3.7.
&
The end of the light cable must not protrude more than 40 mm from the flange of the
Coupler. The diameter of the optic cable at its far end must not exceed 8 mm.