Settling Time - HP E4419B User Manual

Power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

Specifications
Power Meter Supplemental Characteristics
For HP E-series power sensors
In FAST mode, within the range -50 dBm to +17 dBm, for a 10 dB
decreasing power step, the settling time is
1
When a decreasing power step crosses the power sensor's auto-range switch
point, add 25 ms. Refer to the relevant power sensor manual for further
information.
For HP E-series power sensors in normal and x2 speed modes, manual
filter, 10 dB decreasing power step:
Number of
Averages
Settling Time (s)
(Normal Mode)
Response Time (s)
(x2 mode)
5-10
Table 5-5: Settling Time
1
2
4
8
0.07
0.12
0.21
0.4
0.04
0.07
0.12
0.21