Zeroing and Calibrating Using TTL Inputs

You can use the TTL inputs on the rear panel Rmt I/O port to initiate zero
and calibration cycles on the power meter. The connector is an RJ-45
series shielded modular jack with the TTL input pins connected as shown
in Figure 2-3.
Figure 2-3: Rmt I/O Port TTL Inputs
1 2 3 4 5 6 7 8
The TTL inputs are active low and control the zero and calibration
functions as shown in Table 2-2
Table 2-2: TTL Inputs Control Logic
Input 1
1
1
0
0
HP E4419B User's Guide
Get other manuals https://www.bkmanuals.com
Zeroing and Calibrating the Power Meter
Pin Number
1
none
2
Ground
3
Upper Window TTL Output
4
Lower Window TTL Output
5
TTL Input 1
6
TTL Input 2
7
Ground
8
Ground
Input 2
Operation
1
None
0
CAL A
1
ZERO BOTH
0
CAL B
Power Meter Operation
Connection
2-13