Warnings
!
other than Atrial Fibrillation. You should notify your physician if you detect possible changes in your health.
DO NOT sterilize using irradiation, steam or ethylene oxide. Refer to cleaning and disinfection instructions.
Use of cleaning agents other than specified may damage the device.
DO NOT use the ScanWatch 2 if it is damaged. Use of a damaged device could cause patient injury or equipment
failure.
DO NOT take recordings when ScanWatch 2 is in close vicinity to strong electromagnetic fields (e.g., electromag-
netic anti-theft systems, metal detectors).
DO NOT take recordings during a medical procedure (e.g., magnetic resonance imaging, diathermy, lithotripsy, cau-
tery and external defibrillation procedures).
DO NOT take recordings when ScanWatch 2 is outside of the operational temperature range (5°C - 40°C) indicated
in the ScanWatch 2 user manual and humidity range of 20% to 90% relative humidity.
DO NOT use to diagnose heart-related conditions.
DO NOT use with a cardiac pacemaker, ICDs, or other implanted electronic devices.
DO NOT take a recording during physical activity.
No modification of this equipment is allowed.
DO NOT modify this equipment without authorization of the manufacturer.
As the power cord of the charger can cause strangulation due to excessive length, keep it away from children
and pets.
If this equipment is modified, appropriate inspection and testing must be conducted to ensure continued safe use
of the equipment. The battery inside the watch will stop charging when the temperature is less than -10°C (+/-5°C
or ) or over 50°C (+/-5°C).