NOTE:
Patient Position:
1) Comfortably seated
2) Legs uncrossed
3) Feet flat on the floor
4) Position with back and arm supported
5) Middle of the CUFF at the level of the right atrium of the heart
Picture-06
5.4. Irregular Heartbeat Detector
This symbol - indicates that certain pulse irregularities were detected
during the measurement.
In this case, the result may deviate from your normal blood pressure – repeat
the measurement.
Consult your physician on frequent appearance of the Irregular Heartbeat
Symbol.
This instrument is an oscillometric blood pressure monitor device that also
analyzes pulse frequency during measurement. The instrument is clinically
tested.
If pulse irregularities occur during measurement, the irregular heartbeat
symbol is displayed after the measurement. If the symbol appears more
frequently (e.g. several times per week on measurements performed daily) or
if it suddenly appears more often than usual, we recommend the patient to
seek medical advice. The instrument does not replace a cardiac examination,
but serves to detect pulse irregularities at an early stage.
12
5.5. Error Messages
The following message will appear on the display indicating:
SYMBOL
CAUSE
No display
Weak battery or improper
appears
placement
Er 1
Sensor abnormality
Monitor could not detect pulse
Er 2
wave or cannot calculate the
blood pressure data
Measurement result is abnormal
Er 3
(SYS≦35mmHg, DIA≦23mmHg)
Too loose cuff or air leakage
(Cannot inflate to 30mmHg within
Er 4
15s)
Er 5
The air tube is crimped
The sensor is sensing great
Er 6
fluctuation in the pressure
The pressure that the sensor
Er 7
sensing is over the limit
The calibration is incorrect or the
Er 8
device has not been calibrated
CORRECTION
Replace both batteries with new ones.
Check the battery installation for proper
placement of the battery polarities.
Check if the pump is working or not. If it
is working, then the problem is sensor
abnormality. Please contact customer
service for advice.
Check if the air releasing is too slow or not.
If it is too slow, please check if there is any
dust in the tube plug of the cuff and the cuff
port in the device. If yes, please clean and
start the measurement again. If no, please
contact customer service for advice.
Occurs once. Please repeat measurement.
Occurs continuously, contact Medescan
customer service
Readjust the cuff correctly and make sure
the air plug is properly inserted in the unit
Correct it and take the measurement again
Please keep quiet and don't move
Please contact Medescan Customer service
Please contact Medescan Customer service